Paradigm REAL-Time REVEL Insulin Infusion Pump MMT-723LNAB: medical device reports filed with FDA

1,889 reports name it, 2014–2026. Manufacturer given most often on reports: Medtronic Puerto Rico Operations. Product code OYC.

1,889
device reports naming the brand
0% of all MAUDE reports · about 156 a year
11
reports, 12 months to August 2026
6 in the 12 months before
60.8%
classified as malfunction
85.5% across the product code
0.8%
classified as death, as reported
15 reports · not verified by FDA

FDA's MAUDE database holds 1,889 reports that name the brand "Paradigm REAL-Time REVEL Insulin Infusion Pump MMT-723LNAB" (pump infusion insulin to be used with invasive glucose sensor), received between June 2014 and August 2026; the manufacturer given most often on reports is Medtronic Puerto Rico Operations. Spellings of one product vary from report to report.

11 reports arrived in the 12 months to August 2026, up 83% from 6 in the 12 months before. FDA releases MAUDE data monthly; reports can arrive months after an event, so the newest months grow in later releases. The event types recorded are malfunction (60.8%), injury (38.4%) and death (0.8%); across all pump, infusion, insulin, to be used with invasive glucose sensor reports (product code OYC) death is recorded in 0.3% and malfunction in 85.5%. Event type is the reporter's classification of the report, not a verified outcome.

The device problems coded most often are device displays incorrect message (14.6%), mechanical problem (13.9%) and device difficult to program or calibrate (13.1%). The patient problems coded most often are hyperglycemia, hypoglycemia and elevated ketones/ diabetic ketoacidosis. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data."

Reports by month, five years

01428Sep 2021: 7Oct 2021: 4Nov 2021: 3Dec 2021: 32022Jan 2022: 3Feb 2022: 6Mar 2022: 28Apr 2022: 6May 2022: 3Jun 2022: 1Jul 2022: 9Aug 2022: 1Sep 2022: 5Oct 2022: 0Nov 2022: 4Dec 2022: 42023Jan 2023: 1Feb 2023: 6Mar 2023: 4Apr 2023: 1May 2023: 0Jun 2023: 0Jul 2023: 0Aug 2023: 0Sep 2023: 2Oct 2023: 2Nov 2023: 1Dec 2023: 12024Jan 2024: 1Feb 2024: 0Mar 2024: 0Apr 2024: 1May 2024: 2Jun 2024: 1Jul 2024: 1Aug 2024: 1Sep 2024: 0Oct 2024: 0Nov 2024: 0Dec 2024: 02025Jan 2025: 1Feb 2025: 0Mar 2025: 0Apr 2025: 0May 2025: 0Jun 2025: 1Jul 2025: 3Aug 2025: 1Sep 2025: 3Oct 2025: 1Nov 2025: 0Dec 2025: 22026Jan 2026: 0Feb 2026: 1Mar 2026: 0Apr 2026: 2May 2026: 0Jun 2026: 0Jul 2026: 0Aug 2026: 2

Reports by the month FDA received them. FDA releases MAUDE data monthly; reports can arrive months after an event, so the newest months grow in later releases.

By year received

03096182014: 320142015: 32017: 44720172018: 6182019: 45720192020: 1752021: 7420212022: 702023: 1820232024: 72025: 1220252026: 5

Event type and report source

Event type, as classified on the report

Death0.8%15
Injury38.4%726
Malfunction60.8%1,148
Other0%0
Not given0%0

Who filed the report

Manufacturer report100%1,889
Voluntary report0%0
User facility report0%0
Distributor report0%0
Not given0%0

Grey bar: all 891,173 reports carrying product code OYC. Event type is the filer's classification and is not verified by FDA.

Device problems coded

Device problem (FDA code)ReportsShareShare across product code
Device displays incorrect message27514.6%24%
Mechanical problema mechanical problem26213.9%3.7%
Device difficult to program or calibrate24713.1%2.8%
Obstruction of flow1809.5%3%
Patient device interaction problem1769.3%7.9%
No display/image1759.3%5%
Power problem1477.8%4.9%
Battery problema problem with the battery844.4%3.1%
Breakthe device broke834.4%10.1%
Display or visual feedback problem663.5%0.9%
Insufficient device problem information613.2%2.1%
Device operates differently than expectedthe device behaved unexpectedly573%4.5%
Priming problem573%0.4%
Device alarm system542.9%2.4%
Premature discharge of battery532.8%0.8%
Device inoperablethe device could not be used492.6%2.7%
Failure to deliverthe device did not deliver the dose or item492.6%1.4%
Detachment of device or device componentpart of the device came off402.1%0.6%
Failure to cycle321.7%0.7%
Insufficient flow or under infusion271.4%0.7%

FDA's device problem codes as entered on the report; one report can carry several. Generic codes such as "no apparent adverse event" are left out.

Patient problems coded

Patient problem (FDA code)ReportsShare
Hyperglycemiahigh blood sugar67435.7%
Hypoglycemialow blood sugar1799.5%
Elevated ketones/ diabetic ketoacidosis331.7%
Deaththe patient died; cause not stated by this term160.8%
Loss of consciousnesspassing out80.4%
Seizures60.3%
Coma50.3%
Vomitingbeing sick50.3%
Nauseafeeling sick40.2%
Blood loss30.2%
Chest painchest pain30.2%
Painpain, site not specified30.2%
Purulent discharge20.1%
Shock, insulin20.1%
Skin irritationirritated skin20.1%

Patient problem codes are present mainly in reports filed since 2019 and are often blank.

In context

MeasureParadigm REAL-Time REVEL Insulin Infusion Pump MMT-723LNABProduct code OYCAll MAUDE reports
Reports1,889891,17326,136,888
Share of that pool—0.2%0%
Classified as death, per 1,000 reports839
Classified as injury38.4%14.2%36.2%
Classified as malfunction60.8%85.5%62.3%
Filed by the manufacturer100%100%94.9%

Shares are of reports, not of devices in use. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data."

Other brands with product code OYC

BrandReportsLatest 12 monthsClassified as death
Paradigm Real-Time Revel Insulin Infusion Pump178,69100.6%
Paradigm Real-Time Insulin Infusion Pump126,28200.3%
640G Insulin Pump MMT-1712K119,5471,5980.1%
640G Insulin Pump MMT-1711K76,6091510.1%
Animas Vibe88,70300%
T Slim G5 System49,78300%
t:slim G4 system39,215200%
Pump Mmt-1712kl 640g V4 10 Bk Sf Mg21,5319520%
640g Insulin Pump16,91100%
Paradigm REAL-Time REVEL Insulin Infusion Pump MMT-723LNAS12,936410.8%
t:slim X2 Insulin Pump31,159860%
Pump Mmt-1711kl 640g V4 10 Bk Sf Mm11,1571680%
Paradigm REAL-Time Insulin Infusion Pump MMT-754WWS10,4656610%
Paradigm Real-Time Revel Insulin Infusion Pump Mmt-723nas8,540490.8%
640g Insulin Pump Mmt-1712b5,62910%

Ordered by report count for navigation only. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data." FDA also states that "common products may have a higher number of adverse events due to the higher total number of people using the product."

Questions about Paradigm REAL-Time REVEL Insulin Infusion Pump MMT-723LNAB reports

How many FDA reports name Paradigm REAL-Time REVEL Insulin Infusion Pump MMT-723LNAB?

1,889 medical device reports in the openFDA MAUDE release of 28 Sep 2026, covering reports received through August 2026; 11 in the latest 12 months.

What kinds of events are reported?

malfunction (60.8%), injury (38.4%) and death (0.8%). "Event type" is how the report was classified when filed: death, injury, malfunction or other.

What device problems are coded?

device displays incorrect message (14.6%), mechanical problem (13.9%), device difficult to program or calibrate (13.1%), obstruction of flow (9.5%) and patient device interaction problem (9.3%). These are FDA's device problem codes as entered on the report.

Who files these reports?

manufacturer reports (100%). Manufacturers and user facilities are required to report; voluntary reports come from health professionals and the public.

Do these reports show that Paradigm REAL-Time REVEL Insulin Infusion Pump MMT-723LNAB was at fault?

No. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data." FDA adds that "a causal relationship cannot be established between product and reactions listed in a report."

How do I report a problem with a medical device?

Through FDA MedWatch (Form FDA 3500 for health professionals and consumers), or to the manufacturer. See the guide on how to report.

Where this comes from. Counts are built from the openFDA Device Adverse Event (MAUDE) bulk export released 28 Sep 2026, which holds reports FDA received through 31 Aug 2026. Brand, generic and manufacturer names are as written on the reports; product code names come from FDA's device classification. A report counts here when any device it lists carries this brand name after punctuation and capitalisation are normalised. No report narrative, lot number, patient detail or reporter identity is published. See the methodology and sources; FDA's record is authoritative, and corrections are handled within five working days.

Drug data (FAERS) through 30 Jun 2026; device data (MAUDE) through 31 Aug 2026. Not medical advice.