Devices › Dc-defibrillator, low-energy, (including paddles)

Device brand name · Defibrillator

Pd1200 Defibrillator/Pacemaker: medical device reports filed with FDA

1,380 reports name it, 1996–2011. Manufacturer given most often on reports: Zoll Medical. Product code LDD.

1,380
device reports naming the brand
0% of all MAUDE reports · about 46 a year
0
reports, 12 months to August 2026
0 in the 12 months before
94.8%
classified as malfunction
89.2% across the product code
4.9%
classified as death, as reported
67 reports · not verified by FDA

1,380 medical device reports received by FDA name the brand "Pd1200 Defibrillator/Pacemaker" (defibrillator); the manufacturer given most often on reports is Zoll Medical; received from September 1996 to December 2011. Brand names are as the reporter or manufacturer wrote them, so one product can appear under several spellings.

0 reports arrived in the 12 months to August 2026, and none in the 12 months before. FDA releases MAUDE data monthly; reports can arrive months after an event, so the newest months grow in later releases. The event types recorded are malfunction (94.8%), death (4.9%) and injury (0.1%); across all dc-defibrillator, low-energy, (including paddles) reports (product code LDD) death is recorded in 5.7% and malfunction in 89.2%. Event type is the reporter's classification of the report, not a verified outcome.

The device problems coded most often are failure to power up (17.4%), device displays incorrect message (14.2%) and replace (11.2%). The patient problems coded most often are no patient involvement, death and therapy/non-surgical treatment/delayed/prolonged. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data."

Reports by month, five years

No report naming Pd1200 Defibrillator/Pacemaker was received in the five years to August 2026; the latest was received in December 2011.

By year received

01172341996: 3019961997: 851998: 23419981999: 2202000: 14520002001: 1332002: 10320022003: 702004: 4520042005: 692006: 7220062007: 652008: 4120082009: 302010: 2720102011: 11

Event type and report source

Event type, as classified on the report

Death4.9%67
Injury0.1%1
Malfunction94.8%1,308
Other0%0
Not given0.3%4

Who filed the report

Manufacturer report100%1,380
Voluntary report0%0
User facility report0%0
Distributor report0%0
Not given0%0

Grey bar: all 16,581 reports carrying product code LDD. Event type is the filer's classification and is not verified by FDA.

Device problems coded

Device problem (FDA code)ReportsShareShare across product code
Failure to power up24017.4%8.3%
Device displays incorrect message19614.2%12.3%
Replace15511.2%11.5%
Failure to discharge13810%7.7%
Failure to charge1269.1%8.6%
Loss of powerthe device lost power1138.2%4.8%
Other (for use when an appropriate device code cannot be identified)936.7%3%
No display/image674.9%6.3%
Device issuea problem with the device584.2%3.2%
Energy output problem533.8%1.3%
Failure, intermittent533.8%2.6%
Pacing problem513.7%6.6%
Blank screen231.7%1.8%
Pacing inadequately211.5%0.4%
Failure to capture171.2%0.6%
Battery failure151.1%1.1%
Defibrillation/stimulation problem130.9%3.3%
No device output130.9%0.6%
Noise130.9%0.5%
Output above specifications130.9%0.2%

FDA's device problem codes as entered on the report; one report can carry several. Generic codes such as "no apparent adverse event" are left out.

Patient problems coded

Patient problem (FDA code)ReportsShare
No patient involvement27720.1%
Deaththe patient died; cause not stated by this term473.4%
Therapy/non-surgical treatment/delayed/prolonged322.3%
Electric shockan electric shock130.9%
Burn(s)a burn90.7%
Ventricular tachycardiaa fast rhythm from the lower heart chambers90.7%
Atrial fibrillationan irregular heart rhythm70.5%
Therapy/non-surgical treatment, additional60.4%
Ventricular fibrillationa chaotic rhythm from the lower heart chambers60.4%
Asystole50.4%
Other (for use when an appropriate patient code cannot be identified)50.4%
Cardiac arrestthe heart stopped40.3%
Resuscitation40.3%
Bradycardiaslow heart rate30.2%
Burn, thermal30.2%

Patient problem codes are present mainly in reports filed since 2019 and are often blank.

In context

MeasurePd1200 Defibrillator/PacemakerProduct code LDDAll MAUDE reports
Reports1,38016,58126,136,888
Share of that pool—8.3%0%
Classified as death, per 1,000 reports49579
Classified as injury0.1%2.7%36.2%
Classified as malfunction94.8%89.2%62.3%
Filed by the manufacturer100%92.7%94.9%

Shares are of reports, not of devices in use. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data."

Other brands with product code LDD

BrandReportsLatest 12 monthsClassified as death
Lifepak 20 Defibrillator/Monitor2,75310.8%
Lifepak 12 Defibrillator/Monitor Series3,60804.9%
Pd1400 Defibrillator/Pacemaker1,77706.1%
Lifepak R 20 Defibrillator/Monitor2,24840.5%
Lifepak 10 Defibrillator/Monitor/Pacemaker62204.7%
Lifepak 9p Defibrillator/Monitor/Pacemaker62802.9%
Codemaster Xl42702.1%
Autoclavable Internal Handles Mseries37910.5%
Tempus LS-Manual28586.3%
Lifepak 9 Defibrillator/Monitor26405.7%
Heartstart Mrx18,59200.7%
Codemaster Xl+25102.8%
D900 Defibrillator21503.3%
Lifepak 11 Diagnostic Cardiac Monitor21002.4%
Codemaster 10019805.1%

Ordered by report count for navigation only. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data." FDA also states that "common products may have a higher number of adverse events due to the higher total number of people using the product."

Questions about Pd1200 Defibrillator/Pacemaker reports

How many FDA reports name Pd1200 Defibrillator/Pacemaker?

1,380 medical device reports in the openFDA MAUDE release of 28 Sep 2026, covering reports received through August 2026; 0 in the latest 12 months.

What kinds of events are reported?

malfunction (94.8%), death (4.9%) and injury (0.1%). "Event type" is how the report was classified when filed: death, injury, malfunction or other.

What device problems are coded?

failure to power up (17.4%), device displays incorrect message (14.2%), replace (11.2%), failure to discharge (10%) and failure to charge (9.1%). These are FDA's device problem codes as entered on the report.

Who files these reports?

manufacturer reports (100%). Manufacturers and user facilities are required to report; voluntary reports come from health professionals and the public.

Do these reports show that Pd1200 Defibrillator/Pacemaker was at fault?

No. FDA states that "MDR data alone cannot be used to establish rates of events, evaluate a change in event rates over time or compare event rates between devices. The number of reports cannot be interpreted or used in isolation to reach conclusions about the existence, severity, or frequency of problems associated with devices." It adds that the reporting system "has limitations, including the potential submission of incomplete, inaccurate, untimely, unverified, or biased data." FDA adds that "a causal relationship cannot be established between product and reactions listed in a report."

How do I report a problem with a medical device?

Through FDA MedWatch (Form FDA 3500 for health professionals and consumers), or to the manufacturer. See the guide on how to report.

Where this comes from. Counts are built from the openFDA Device Adverse Event (MAUDE) bulk export released 28 Sep 2026, which holds reports FDA received through 31 Aug 2026. Brand, generic and manufacturer names are as written on the reports; product code names come from FDA's device classification. A report counts here when any device it lists carries this brand name after punctuation and capitalisation are normalised. No report narrative, lot number, patient detail or reporter identity is published. See the methodology and sources; FDA's record is authoritative, and corrections are handled within five working days.

Drug data (FAERS) through 30 Jun 2026; device data (MAUDE) through 31 Aug 2026. Not medical advice.